FDA Adverse Event Malfunction Summary report: N

HAMILTON-C6

MDR report key: 19496467 · Received June 10, 2024

Report

Report Number
3001421318-2024-01385
Event Type
Malfunction
Date Received
June 10, 2024
Date of Event
June 4, 2024
Report Date
December 17, 2024
Manufacturer
HAMILTON MEDICAL AG
Product Code
CBK
PMA / PMN Number
K201658
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AG COMES TO THE CONCLUSION: INVESTIGATION ONGOING. HAMILTON MEDICAL AG COMPLAINT NUMBER: (B)(4). FOLLOW-UP 1 - CORRECTED INFORMATION: **UDI RELATED DATA QUALITY UPDATES ONLY** FIELD D4 WAS UPDATED WITH FULL UDI INFORMATION. UPDATED FIELDS: B4, D1, D4, G6, H2, H4, H5, H11.

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AG COMES TO THE CONCLUSION: INVESTIGATION ONGOING. A DETAILED INVESTIGATION WAS PERFORMED BY AN EXPERT FROM THE TECHNICAL SERVICE: THE INVESTIGATION CONFIRMED THAT DEVICE DID NOT WORK AS INTENDED AT THE TIME OF THE EVENT. THE INCIDENT REPORTED BY THE CUSTOMER OCCURRED DURING START-UP WITHOUT PATIENT INVOLVEMENT. THEREFORE, THERE WAS NO PATIENT OR USER HARM. THE ROOT CAUSE WAS A DEFECTIVE FFC-CABLE (PART N° 161554). THE DEFECTIVE CABLE WAS EXCHANGED, THE DEVICE WAS TESTED AND WORKED AS INTENDED AGAIN.

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AG COMES TO THE CONCLUSION: INVESTIGATION ONGOING.

Description of Event or Problem · 0

THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: SUMMARY:TF 243004 (BUZZER DEFECT AT STARTUP) OR BUZZER DEFECTIVE.

Description of Event or Problem · 0

THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: SUMMARY:TF 243004 (BUZZER DEFECT AT STARTUP) OR BUZZER DEFECTIVE.

Description of Event or Problem · 0

THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: SUMMARY:TF 243004 (BUZZER DEFECT AT STARTUP) OR BUZZER DEFECTIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
206207 HAMILTON-C6 HAMILTON-C6 VENTILATOR CBK HAMILTON MEDICAL AG 160021

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown