FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 19495618 · Received June 9, 2024

Report

Report Number
2955842-2024-15349
Event Type
Malfunction
Date Received
June 9, 2024
Date of Event
March 29, 2024
Report Date
May 15, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874112298
PMA / PMN Number
K112263
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) HAS NOT RECEIVED THE INSTRUMENT FOR EVALUATION. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE PRODUCT IS RETURNED AND EVALUATED AND/OR IF ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL (IS) OBTAINED THE FOLLOWING ADDITIONAL INFORMATION REGARDING THIS EVENT: THE INSTRUMENT HAS BEEN SHIPPED.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE MONOPOLAR CURVED SCISSORS INSTRUMENT WAS ANALYZED AND FOUND TO EXHIBIT SCRATCH MARKS/ABRASIONS ON THE BROWN LASER MARKED SECTION OF THE TUBE EXTENSION. AS A RESULT, THE ORANGE PLASTIC BENEATH THE LASER MARKING WAS EXPOSED. TUBE EXTENSION SCRATCH MARKS MEASURED ROUGHLY 2.6MM X 1.0MM. THERE WAS NO MATERIAL MISSING. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED PROSTATECTOMY SURGICAL PROCEDURE, THE DISTAL PART OF THE MONOPOLAR CURVED SCISSORS (MCS) INSULATING SHEATH WAS DAMAGED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT CABLE INSULATION WAS DAMAGED. IT WAS NOT POSSIBLE TO CONFIRM THE TYPE OF DAMAGE OBSERVED.

Description of Event or Problem · 0

REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2125828 ENDOWRIST MONOPOLAR CURVED SCISSORS NAY INTUITIVE SURGICAL, INC 470179-19 K13231130 0090 00886874112298

Patients

Seq Age Sex Outcome Treatment
1 NA Male DA VINCI INSTRUMENTS AND ACCESSORIES.