FDA Adverse Event Malfunction Summary report: N

ALINITY C PROCESSING MODULE

MDR report key: 19495239 · Received June 9, 2024

Report

Report Number
3016438761-2024-00342
Event Type
Malfunction
Date Received
June 9, 2024
Date of Event
May 28, 2024
Report Date
June 19, 2024
Manufacturer
ABBOTT LABORATORIES
Product Code
JJE
UDI-DI
00380740137380
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FIELD SERVICE REPRESENTATIVE (FSR) INSPECTED THE INSTRUMENT AND RESOLVED THE ISSUE BY ADJUSTING THE PUMP, NRD-30 MAG., CUVETTE AND WASH CUPS, N. S. AS THE CUVETTE WASHER FLOW WAS LOW. NO ADDITIONAL DISCREPANT RESULT ISSUES HAVE BEEN REPORTED SINCE THE SERVICE ACTIVITY WAS COMPLETED. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. THE INSTRUMENT SERVICE HISTORY REVIEW FOR (B)(6) REVEALED NO ADDITIONAL SERVICE TICKETS ASSOCIATED WITH DISCREPANT/ERRATIC RESULTS. THERE WERE NO ADDITIONAL SERVICE OR COMPLAINT ISSUES ON OR AROUND THE DATE THIS COMPLAINT WAS INITIATED THAT MAY HAVE CONTRIBUTED TO THIS ISSUE. A REVIEW OF TRACKING AND TRENDING FOR THE ALINITY C PROCESSING MODULE DID NOT IDENTIFY ANY TRENDS. A REVIEW OF TRACKING AND TRENDING FOR THE PUMP, NRD-30 MAG., CUVETTE AND WASH CUPS, N. S. DID NOT IDENTIFY ANY TRENDS. REVIEW OF THE MANUFACTURING DOCUMENTATION DID NOT IDENTIFY ANY NON-CONFORMANCES ASSOCIATED WITH THE COMPLAINT ISSUE. LABELING WAS REVIEWED AND FOUND TO ADEQUATELY ADDRESS THE ISSUE UNDER REVIEW. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY C PROCESSING MODULE FOR SERIAL (B)(6) OR THE PUMP, NRD-30 MAG., CUVETTE AND WASH CUPS, N. S. WERE IDENTIFIED.

Additional Manufacturer Narrative · 0

COMPLETED INFORMATION FOR SECTION A1 PATIENT IDENTIFIER: (B)(6). ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSELY DEPRESSED ALINITY C GLUCOSE RESULT FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) INITIAL RESULT = 13 MG/DL, REPEAT = 123 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSELY DEPRESSED ALINITY C GLUCOSE RESULT FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6). INITIAL RESULT = 13 MG/DL, REPEAT = 123 MG/DL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2125761 ALINITY C PROCESSING MODULE ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE ABBOTT LABORATORIES 00380740137380

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ALNTY C GLUCOSE 400T, 07P55-20, (B)(6) | ALNTY C GLUCOSE 400T, 07P55-20, (B)(6)