FDA Adverse Event Malfunction Summary report: N

GALILEO LAG SCREW, RIGHT, Ø10.5MMX100MM

MDR report key: 19494032 · Received June 8, 2024

Report

Report Number
1220246-2024-05295
Event Type
Malfunction
Date Received
June 8, 2024
Date of Event
May 8, 2024
Report Date
June 11, 2025
Manufacturer
ARTHREX, INC.
Product Code
HSB
UDI-DI
00848665015850
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD NOT BE DETERMINED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION. IF THE DEVICE BECOMES AVAILABLE FOR EVALUATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: G3, H3, H6 BASED ON THE INFORMATION PROVIDED WHICH MAY INCLUDE THE DEVICE (IF AVAILABLE AND RETURNED), PICTURES, VIDEOS, EVENT DESCRIPTION, AND ANY ADDITIONAL INFORMATION FROM THE FIELD, ARTHREX WAS ABLE TO CONCLUDE A MOST LIKELY CAUSE. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO USER ERROR DUE TO MISALIGNED INSERTION AND/OR IMPROPER BONE PREPARATION.

Description of Event or Problem · 0

ON 05/09/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN 1192-100 TELESCOPING LAG SCREW, A 1099-110 TELESCOPING LAG SCREW, AND A 1099-100 TELESCOPING LAG SCREW WOULD NOT ADVANCE INTO THE FEMORAL HEAD OF AN AR-9094-10-2030 TROCHANTERIC NAIL. THE TELESCOPING LAG SCREWS WERE USED WITH A NEW TROCHANTERIC NAIL; HOWEVER, THE ISSUE PERSISTED, AND THE TELESCOPING LAG SCREWS WOULD STILL NOT ADVANCE. THE CASE WAS COMPLETED USING A NEW TROCHANTERIC NAIL AND A NEW 1099-110 TELESCOPING LAG SCREW. THE CASE WAS DELAYED ABOUT FIFTEEN MINUTES. THIS WAS DISCOVERED DURING A LEFT HIP IM NAIL PROCEDURE ON (B)(6) 2024.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2109802 GALILEO LAG SCREW, RIGHT, Ø10.5MMX100MM INTRAMED FIXATN ROD AND ACCESS HSB ARTHREX, INC. GALILEO LAG SCREW, RIGHT, Ø10.5MMX100MM 210370 00848665015850

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown