FDA Adverse Event Malfunction Summary report: N

AUTOSOFT 90

MDR report key: 19492068 · Received June 7, 2024

Report

Report Number
3003442380-2024-07011
Event Type
Malfunction
Date Received
June 7, 2024
Date of Event
March 27, 2024
Report Date
June 7, 2024
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244018334
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR 1882064 - DEVICE 5 OF 12. PATIENT CITY: (B)(6).

Description of Event or Problem · 0

UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED TWELVE INFUSION SET CANNULA BENT EVENTS FROM (B)(6) 2024 TO (B)(6) 2024. ALL THE EVENTS OCCURRED WITHIN 3 HOURS OF INSERTION AND THE INSERTION SITE WAS AT ABDOMEN AND THIGH. THE BLOOD GLUCOSE LEVEL AT THE TIME OF ALL EVENTS WAS BETWEEN 200 TO 400MG/DL AND PATIENT RESOLVED IT BY CHANGING SOFT CANNULA INFUSION SET AND RESUMED INSULIN SUCCESSFULLY. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2319766 AUTOSOFT 90 UNO INSET II 60/9 BLUE TCAP 10PK INT FPA UNOMEDICAL A/S 1002824 6003405 05705244018334

Patients

Seq Age Sex Outcome Treatment
1 55 YR Male