FDA Adverse Event Malfunction Summary report: N

ADVANCED PERFUSION SYSTEM 1

MDR report key: 19490159 · Received June 7, 2024

Report

Report Number
1828100-2024-00154
Event Type
Malfunction
Date Received
June 7, 2024
Date of Event
May 14, 2024
Report Date
August 15, 2024
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
Product Code
DTQ
UDI-DI
00886799000588
PMA / PMN Number
K172220
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PLEASE DISREGARD THE PREVIOUSLY SUBMITTED MEDWATCHES RELATED TO THIS COMPLAINT. THERE WAS NO MALFUNCTION IN THE INTENDED PERFORMANCE ASSOCIATED WITH THE SERVICE MESSAGE AS THE DEVICE PERFORMED TO SPECIFICATIONS. DURING LABORATORY ANALYSIS, THE PRODUCT SURVEILLANCE TECHNICIAN (PST) CONNECTED THE ELECTRONIC PATIENT GAS SYSTEM (EPGS) TO LAB USE ONLY (LUO) TESTING EQUIPMENT. UPON POWER UP OF THE EPGS PERFUSION SCREEN, THE PST OBSERVED A 'SERVICE GAS SYSTEM' MESSAGE AND CALIBRATION WAS UNAVAILABLE. IN THE SERVICE SCREEN, IT WAS NOTED THAT THE OXYGEN (O2) PERCENT HOURS READ 2211802. AFTER 2000000% HOURS, THE SYSTEM ALERTS THE USER TO 'SERVICE GAS SYSTEM'. THE UNIT OPERATED AS INTENDED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Additional Manufacturer Narrative · 0

EVALUATION IS IN PROGRESS, BUT NOT YET CONCLUDED.

Description of Event or Problem · 0

IT WAS THAT DURING PRIMING OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE CENTRAL CONTROL MONITOR (CCM) DISPLAYED A 'GAS BLENDER FAILURE' MESSAGE. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2125190 ADVANCED PERFUSION SYSTEM 1 CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 DTQ TERUMO CARDIOVASCULAR SYSTEMS CORPORATION 801188 00886799000588

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown