AUTOSOFT XC
Report
- Report Number
- 3003442380-2024-07021
- Event Type
- Malfunction
- Date Received
- June 7, 2024
- Date of Event
- July 23, 2023
- Report Date
- June 7, 2024
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244016651
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR 1881862-DEVICE 4 OF 10
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED 10 INFUSION SET CANNULA BENT EVENT ON (B)(6) 2023 AFTER 3 OR MORE HOURS AFTER INSERTION. INSERTION SITE WAS AT ABDOMEN. CUSTOMER REGULARLY ROTATE SITE LOCATION. INSERTION SET HAS BEEN USED MORE THAN 3 HOURS BUT LESS THAN 6 HOURS. THE BLOOD GLUCOSE LEVEL WAS 300-400MG/DL. THE CUSTOMER REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2107241 | AUTOSOFT XC | UNO INSET I 60/9 GREY TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1001681 | UNKNOWN | 05705244016651 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Female |