AUTOSOFT XC
Report
- Report Number
- 3003442380-2024-07128
- Event Type
- Malfunction
- Date Received
- June 7, 2024
- Date of Event
- April 19, 2024
- Report Date
- June 7, 2024
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244021099
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR 1882091. DEVICE 4 OF 4.
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED SIMILAR EVENTS OF BENT CANNULA WITH FOUR INFUSION SETS. THE SYMPTOMS WERE NOTICED WITHIN 3 HOURS OF INSERTION. THE SITE OF INSERTION WAS ABDOMEN AND WAS ROTATED REGULARLY. PATIENT'S BLOOD GLUCOSE AT THE TIME OF EVENT WAS 302 MG/DL. THEREFORE, PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2107215 | AUTOSOFT XC | UNO INSET I 80/6 GREY TCAP 10PACK US | FPA | UNOMEDICAL A/S | 1003912 | 6004706 | 05705244021099 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Male |