FDA Adverse Event Injury Summary report: N

EVIS EXERA II DUODENOVIDEOSCOPE

MDR report key: 19481357 · Received June 6, 2024

Report

Report Number
2429304-2024-0000290
Event Type
Injury
Date Received
June 6, 2024
Date of Event
March 7, 2022
Report Date
June 5, 2024
Manufacturer
AIZU OLYMPUS CO., LTD.
Product Code
FDT
UDI-DI
04953170367311
Adverse Event
Yes
Product Problem
Yes
Report Source
Distributor report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Description of Event or Problem · 0

OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED "EVIDENCE OF TRANSMISSION OF NEW DELHI METALLO- LACTAMASEPRODUCING KLEBSIELLA PNEUMONIAE THROUGH A GASTROINTESTINAL ENDOSCOPE WITHOUT AN ELEVATOR CHANNEL" ABSTRACT OBJECTIVE: TO INVESTIGATE THE SOURCE AND TRANSMISSION DYNAMICS OF AN ENDOSCOPE-ASSOCIATED NEW DELHI METALLO--LACTAMASE-PRODUCING KLEBSIELLA PNEUMONIA (NDM-KP) OUTBREAK. DESIGN: EPIDEMIOLOGICAL AND GENOMIC INVESTIGATION. SETTING: (B)(6) HOSPITAL . PATIENTS: FIVE PATIENTS WITH ACTIVE NDM-KP INFECTION IDENTIFIED ON CLINICAL ISOLATES, AND FOUR NDM-KP COLONIZED PATIENTS IDENTIFIED VIA RECTAL SWAB SCREENING. RESULTS: OVER A TWELVE-MONTH PERIOD, NINE PATIENTS WERE IDENTIFIED WITH NDM-KP INFECTION OR COLONIZATION. WHOLE-GENOME SEQUENCING (WGS) REVEALED THAT ALL OF THE IDENTIFIED CASES WERE RELATED BY 25 MUTATIONAL EVENTS OR LESS. SEVEN OF THE CASES WERE LINKED TO GASTROINTESTINAL ENDOSCOPIC PROCEDURES (FOUR CLINICAL CASES AND THREE POSITIVE SCREENS AMONG PATIENTS EXPOSED TO ENDOSCOPES SUSPECTED OF TRANSMISSION). TWO CASES DEMONSTRATED DELAYED TRANSMISSION THAT OCCURRED FIVE MONTHS AFTER THE INITIAL OUTBREAK, LIKELY THROUGH SHARED USAGE OF A NON-THERAPEUTIC GASTROSCOPE WITHOUT AN ELEVATOR CHANNEL. CONCLUSIONS: ALTHOUGH ALL ENDOSCOPE CULTURES IN OUR INVESTIGATION WERE NEGATIVE, THE EPIDEMIOLOGICAL LINK TO GASTROINTESTINAL ENDOSCOPES, THE HIGH DEGREE OF RELATEDNESS VIA WGS, AND THE IDENTIFICATION OF ASYMPTOMATIC NDM-KP COLONIZATION AMONG PATIENTS EXPOSED TO SHARED ENDOSCOPES MAKE THE ENDOSCOPIC MODE OF TRANSMISSION MOST LIKELY. THIS INVESTIGATION HIGHLIGHTS THE PROBABLE TRANSMISSION OF NDM-KP VIA A GASTROSCOPE WITHOUT AN ELEVATOR CHANNEL, OBSERVED SEVERAL MONTHS AFTER AN INITIAL OUTBREAK. WE HYPOTHESIZE THAT PERSISTENT MECHANICAL DEFECTS MAY HAVE CONTRIBUTED TO THE DELAYED DEVICE-RELATED TRANSMISSION OF NDM-KP. PATIENT CLINICAL CULTURE C3 POSITIVE CULTURE FOR NEW DELHI METALLO--LACTAMASE-PRODUCING KLEBSIELLA PNEUMONIAE (NDM-KP) SOURCE OF CULTURE - ABDOMINAL ABSCESS, BLOOD

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1151214 EVIS EXERA II DUODENOVIDEOSCOPE DUODENOVIDEOSCOPE FDT AIZU OLYMPUS CO., LTD. TJF-Q180V 04953170367311

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other