VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE
Report
- Report Number
- 9610773-2024-30969
- Event Type
- Malfunction
- Date Received
- June 6, 2024
- Date of Event
- May 21, 2024
- Report Date
- December 10, 2024
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- HET
- UDI-DI
- 04042761075015
- PMA / PMN Number
- K190744
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
E2/E3: NO INFORMATION AVAILABLE FOR THE OCCUPATION OF THE INITIAL REPORTER OR WHETHER THEY ARE A HEALTHCARE PROFESSIONAL. THE EVALUATION OF THE EVENT IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE DEVICE EVALUATION AND THE LEGAL MANUFACTURER'S FINAL INVESTIGATION AND DUE TO CORRECTIONS. UPDATED FIELDS: B5, D8, D9, E1 (PHONE NUMBER), G3, G6, H2, H3, H4, H6, H11. CORRECTED FIELDS: D3 AND G1 (CORRECTED GERMAN CHARACTER), G4 - PMA/510(K) # ADDED CORRECT DIGITS. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: THE PROTECTOR OF THE VIDEO CABLE SECTION WAS CUT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
IT WAS REPORTED THAT DURING PREPARATION THE RIGID VIDEO SCOPE HAD TWO NOTCHES IN THE CABLE. THERE WERE NO REPORTS OF PATIENT HARM.
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 8158 | VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE | RIGID VIDEO LAPAROSCOPE | HET | OLYMPUS WINTER & IBE GMBH | WA50052A | 04042761075015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |