AUTOSOFT 90
Report
- Report Number
- 3003442380-2024-06292
- Event Type
- Malfunction
- Date Received
- June 6, 2024
- Date of Event
- April 22, 2024
- Report Date
- June 6, 2024
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244014541
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR 1880704 - MDR 3003442380-2024-06292 - DEVICE 1 OF 2. E1: PATIENT CITY: (B)(6).
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED TWO INFUSION SET CANNULA BENT EVENTS ON (B)(6) 2024. BOTH THE EVENT OCCURRED WITHIN 3 HOURS OF INSERTION AND THE INSERTION SITE WAS AT ABDOMEN. THE BLOOD GLUCOSE LEVEL AT THE TIME OF EVENT WAS 494MG/DL AND PATIENT TREATED IT WITH CORRECTION INJECTION VIA MULTIPLE DAILY INJECTION (MDI) FOLLOWED BY CHANGING SOFT CANNULA INFUSION SET AND RESUMED INSULIN SUCCESSFULLY. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1862800 | AUTOSOFT 90 | INSET II 2-PACK 60/6 GREY TCAP | FPA | UNOMEDICAL A/S | 1000281 | UNKNOWN | 05705244014541 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR | Male |