CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS
Report
- Report Number
- 2015691-2024-04348
- Event Type
- Injury
- Date Received
- June 5, 2024
- Date of Event
- May 14, 2024
- Report Date
- July 24, 2024
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- DYE
- UDI-DI
- 00690103176148
- PMA / PMN Number
- P860057
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
H10: ADDITIONAL MANUFACTURER NARRATIVE: THE INVESTIGATION IS STILL IN PROGRESS; THEREFORE, A CONCLUSION HAS YET TO BE ESTABLISHED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY UPON INVESTIGATION COMPLETION. EDWARDS WILL CONTINUE TO REVIEW AND MONITOR ALL REPORTED EVENTS. TRENDS ARE MONITORED ON A MONTHLY BASIS AND IF ACTION IS REQUIRED, APPROPRIATE INVESTIGATION WILL BE PERFORMED.
H11: ADDITIONAL MANUFACTURER NARRATIVE: UPDATED SECTIONS: B4, G3, G6, H2, H6 (INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS). A DEVICE HISTORY RECORD (DHR) REVIEW WAS NOT PERFORMED, AS NO INFORMATION REGARDING A DEVICE FAILURE MODE WAS PROVIDED. FURTHERMORE, THERE IS NO ALLEGATION OF A MALFUNCTION WHICH COULD BE RELATED TO A MANUFACTURING NON-CONFORMANCE AND/OR ONE WAS NOT SUSPECTED OR CONFIRMED THROUGH INVESTIGATION; NO LABELING NON-CONFORMANCE/DEFICIENCY; NO DEVICE-RELATED INFECTION; AND NO EVIDENCE OF A PRODUCT FAILURE WITH REGARD TO DESIGN, RELIABILITY, OR USE ERROR (DOC-0101337). THE COMPLAINT INVESTIGATION WAS CONDUCTED, USING EVALUATION SUMMARY DOC-0212705 AND TRIAGE LEVEL 1, IN ACCORDANCE WITH TCA DOC-0226303 REV. E. DUE DILIGENCE ATTEMPTS WERE MADE TO GATHER ADDITIONAL INFORMATION REGARDING A DEVICE FAILURE MODE, HOWEVER ADDITIONAL INFORMATION IS NOT AVAILABLE AT THIS TIME. PATIENT MEDICAL RECORDS WERE NOT PROVIDED BY THE HOSPITAL FOR REVIEW, OR THE MEDICAL RECORDS PROVIDED DO NOT CLARIFY THE DEVICE FAILURE MODE. PER THE INSTRUCTIONS FOR USE (IFU), REOPERATION/EXPLANT ARE KNOWN POTENTIAL ADVERSE EVENTS ASSOCIATED WITH BIOPROSTHETIC CARDIAC VALVES AND ANNULOPLASTY RINGS. IN THIS CASE, THERE IS NO EVIDENCE OF A MALFUNCTION WHICH COULD BE RELATED TO A MANUFACTURING NON-CONFORMANCE AND/OR ONE WAS NOT SUSPECTED OR CONFIRMED THROUGH INVESTIGATION; NO LABELING NON-CONFORMANCE/DEFICIENCY; NO USE-RELATED ISSUE WITH A HAZARDOUS SITUATION; NO DEVICE-RELATED INFECTION; AND NO EVIDENCE OF A PRODUCT FAILURE WITH REGARD TO DESIGN, RELIABILITY, OR USE ERROR. AS SUCH, NEITHER A PRODUCT RISK ASSESSMENT (PRA) NOR CORRECTIVE OR PREVENTIVE ACTIONS (CAPA/SCAR) ARE REQUIRED AT THIS TIME, PER DOC-0101337 AND DOC-0076862. BASED ON THE INFORMATION AVAILABLE, A DEFINITIVE ROOT CAUSE CANNOT BE CONCLUSIVELY DETERMINED. ALL PERTINENT INFORMATION AVAILABLE TO EDWARDS LIFESCIENCES HAS BEEN SUBMITTED.
IT WAS REPORTED THAT A PATIENT WITH A 23MM 3300TFX VALVE, IMPLANTED IN AORTIC POSITION, UNDERWENT A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF 7 YEARS AND 9 MONTHS DUE TO UNKNOWN REASONS. THE VALVE WAS REPLACED WITH A 23MM 9755RSL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1089517 | CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS | REPLACEMENT HEART-VALVE | DYE | EDWARDS LIFESCIENCES | 3300TFX | 00690103176148 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Female | Hospitalization| R| L |