FDA Adverse Event Malfunction Summary report: N

BD CONNECTA PLUS3 WHITE PEGS

MDR report key: 19472127 · Received June 5, 2024

Report

Report Number
9610847-2024-00148
Event Type
Malfunction
Date Received
June 5, 2024
Date of Event
May 8, 2024
Report Date
July 31, 2024
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMG
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, SA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

OUR QUALITY ENGINEER INSPECTED THE 4 PHOTOS AND 4 SAMPLES SUBMITTED FOR EVALUATION. THE REPORTED ISSUE OF FOREIGN MATTER WAS CONFIRMED UPON INSPECTION AND TESTING OF THE SAMPLES. ANALYSIS OF THE SAMPLE SHOWED THAT THERE WAS EMBEDDED BLACK SPOTS WITHIN THE WHITE MOLDED TAP. BD DETERMINED THAT THE CAUSE OF THE FAILURE WAS RELATED TO OUR MANUFACTURING PROCESS. THE BLACK SPORTS ARE MOST LIKELY BURNT RESIN. EMBEDDED FOREIGN MATTER MAY OCCUR DURING THE MOLDING PROCESS. BURNT EMBEDDED RESIN RESULT FROM MATERIAL BUILD UP IN THE BARREL/SCREW. AS THE MATERIAL FLOWS THROUGH THE BARREL/SCREW THE BUILDUP BREAKS FREE AND ENDS UP IN THE MOLD. MOLDS ARE BEING PURGED BETWEEN THE LOTS TO AVOID THE BUILDUP OF THE BURNT/LOOSE MATERIAL. A REVIEW OF OUR RISK MANAGEMENT DOCUMENTATION WAS PERFORMED AND INDICATES THAT THE POTENTIAL RISK OF THE REPORTED EVENT WAS ASSESSED APPROPRIATELY. PRODUCTION RECORDS WERE REVIEWED, AND THIS BATCH MEETS OUR MANUFACTURING SPECIFICATION REQUIREMENTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD CONNECTA PLUS3 WHITE PEGS HAD FOREIGN MATTER ITEMS CONTAMINATED THE NEWLY OPENED ITEM FOUND CONTAMINATED UNDER TWO LOT NUMBERS 3180404 D & 3032996 C.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1294662 BD CONNECTA PLUS3 WHITE PEGS STOPCOCK, I.V. SET FMG BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 3032996

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown