AUTOSOFT XC
Report
- Report Number
- 3003442380-2024-05867
- Event Type
- Malfunction
- Date Received
- June 5, 2024
- Date of Event
- April 18, 2024
- Report Date
- June 5, 2024
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244016620
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
DEVICE 2 OF 2. E1: PATIENT CITY: (B)(6).
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED TWO INFUSION SET CANNULA BENT EVENTS ON 18-APR-2024. BOTH THE EVENTS OCCURRED WITHIN 3 HOURS OF INSERTION AND THE INSERTION SITE WAS AT ABDOMEN. THE BLOOD GLUCOSE LEVEL AT THE TIME OF SECOND EVENT WAS 170-329 MG/DL. THE PATIENT RESOLVED BOTH THE EVENTS BY CHANGING SOFT CANNULA INFUSION SET AND RESUMED INSULIN SUCCESSFULLY. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1525095 | AUTOSOFT XC | UNO INSET I 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1001680 | 6003941 | 05705244016620 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Female |