CAREASSIST ES120/220/420
Report
- Report Number
- 3006697241-2024-00046
- Event Type
- Malfunction
- Date Received
- June 4, 2024
- Date of Event
- May 30, 2024
- Report Date
- June 4, 2024
- Manufacturer
- HILL-ROM MEXICO
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER
Narratives
THE BAXTER TECHNICIAN FOUND THE POWER CORD NEEDED TO BE REPLACED. PER THE HILLROM SERVICE MANUAL THE BED SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. AN ANNUAL SERVICE OF THE BED IS ADVISED IN ORDER TO MAINTAIN ITS CHARACTERISTICS AND PERFORMANCE. EXAMINE THE PLUG FOR DAMAGE. MAKE SURE THE PLUG IS A ONE-PIECE MOLDED PLUG ASSEMBLY. IF IT IS NOT, REPLACE THE PLUG CORD ASSEMBLY. REPLACE ANY PLUG CORD ASSEMBLY THAT SHOWS ANY OF THESE: ¿ DISCOLORATION OF THE PLUG MOLDING AROUND THE PLUG BLADES; THIS COULD OCCUR IF THE PLUG BLADES HAVE OVERHEATED OR ARCED. ¿ ANY SIGNS OF CRACKING; THIS COULD OCCUR IF THE PLUG HAS BEEN BENT AND STRAIGHTENED TO A POINT PAST ITS USEFUL LIFE. ¿ LOOSE FIT OF THE PLUG BLADE (THE PLUG BLADE MOVES IN THE MOLDING); THIS COULD OCCUR IF THE MOLDING HAS OVERHEATED OR THE BLADES HAVE BEEN BENT AND STRAIGHTENED TO A POINT PAST THEIR USEFUL LIFE. ¿ ANY SIGNS OF DAMAGE TO THE CABLE. ¿ ANY EXPOSED WIRES. REPLACE THE POWER CORD, IF DAMAGED. A SEARCH OF THE BAXTER MAINTENANCE RECORDS SHOWED BAXTER PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN APR 1, 2024. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE POWER CORD TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 174975 | CAREASSIST ES120/220/420 | BED, AC-POWERED ADJUSTABLE HOSPITAL | FNL | HILL-ROM MEXICO | P1170G0000297 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |