FDA Adverse Event Malfunction Summary report: N

MIDMARK

MDR report key: 1946371 · Received December 10, 2010

Report

Report Number
1523530-2010-00014
Event Type
Malfunction
Date Received
December 10, 2010
Date of Event
November 19, 2010
Report Date
December 9, 2010
Manufacturer
MIDMARK CORP.
Product Code
FLE
PMA / PMN Number
K990189
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

WE HAVE REQUESTED THAT THE UNIT BE RETURNED FOR EVALUATION. AS OF THE DATE OF THIS REPORT, THE UNIT HAS NOT BEEN RECEIVED. FROM PHOTOGRAPHS OF THE PC BOARD, IT APPEARS THAT A SOLDER JOINT/RUN ON THE PRINTED CIRCUIT BOARD OVERHEATED CAUSING THE BOARD TO PARTIALLY MELT. THE MELTING PRODUCED THE SMOKE REPORTED BY THE USER. THE ROOT CAUSE OF THE FAILURE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

MAX FROM NASHVILLE CALLED IN, AND SAID THAT THE PCB CAUGHT FIRE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MIDMARK STEAM STERILIZER FLE MIDMARK CORP. M11

Patients

Seq Age Sex Outcome Treatment
1