Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE BRAKE CABLE SLIPPED OUT FROM THE BEARINGS DUE TO OLD BEARINGS. THE TECH REPLACED BOTH BRAKE BEARINGS TO REPAIR THE BED.

Description of Event or Problem · 1

DURING A SAFETY CHECK, THE TECH FOUND THE BRAKE CASTERS WERE NOT HOLDING PROPERLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTRA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 850

Patients

Seq Age Sex Outcome Treatment
1