FDA Adverse Event Injury Summary report: N

VITALITY

MDR report key: 1946261 · Received January 4, 2011

Report

Report Number
2124215-2010-22316
Event Type
Injury
Date Received
January 4, 2011
Date of Event
October 15, 2010
Report Date
October 15, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Removal / Correction Number
Z-1192-06 THRU Z-1194-06
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DECLARED END OF LIFE (EOL) AFTER APPROXIMATELY 55 MONTHS OF IMPLANT, DUE TO MID-LIFE CHARGE TIMES IN EXCESS OF 30 SECONDS. A MANUAL CAPACITOR REFORMATION WAS ATTEMPTED, BUT CHARGE TIMES REMAINED LONG AT 45 SECONDS, CAUSING A DEVICE FAULT CODE TO OCCUR. SUBSEQUENTLY, THIS DEVICE WAS EXPLANTED AND RETURNED TO BOSTON SCIENTIFIC FOR ANALYSIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED AS A RESULT OF THESE OBSERVATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T180

Patients

Seq Age Sex Outcome Treatment
1 45 YR Other| R 6949| T180