MDR report key: 1946173 · Received January 4, 2011

Report

Report Number
2124215-2010-20403
Event Type
Injury
Date Received
January 4, 2011
Date of Event
October 17, 2010
Report Date
June 30, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN ATTEMPT HAS BEEN MADE TO GATHER ADDITIONAL INFORMATION ABOUT THIS EVENT. THIS EVENT WILL BE UPDATED IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. THE CLINICAL OBSERVATION WAS NOT ABLE TO BE CONFIRMED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) PATIENT WAS IN PAIN AND FEELING DIZZY. THE PATIENT REPORTEDLY BLACKED OUT AND FELL. A PATIENT ADVOCATE DISCUSSED HAVING THE PATIENT PERFORM A LATITUDE REMOTE INTERROGATION TO CHECK THE DEVICE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENT DIED OF UNKNOWN CAUSES. THERE WERE NO ALLEGATIONS OR COMPLAINTS AGAINST THE DEVICE'S OPERATION OR FUNCTIONALITY, AND NO REASONABLE EVIDENCE TO INDICATE THE FUNCTION OF THE DEVICE CAUSED OR CONTRIBUTED TO THE PATIENT'S DEATH. THE DEVICE WAS EXPLANTED AND HAS BEEN RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND E110

Patients

Seq Age Sex Outcome Treatment
1 79 YR Other 0184| E110| 4469