FINELINE II
MDR report key: 1946094
·
Received January 4, 2011
Report
- Report Number
- 2124215-2010-20312
- Event Type
- Injury
- Date Received
- January 4, 2011
- Date of Event
- October 14, 2010
- Report Date
- October 14, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- PMA / PMN Number
- P960004
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THIS LEAD WAS SURGICALLY ABANDONED AND A NEW LEAD WAS SUCCESSFULLY IMPLANTED. THERE IS NO FURTHER INFORMATION AVAILABLE AT THIS TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE TO OUR COMPANY, THIS EVENT WOULD BE UPDATED. THE INVESTIGATION IS COMPLETE AT THIS TIME.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED IMPEDANCE MEASUREMENTS GREATER THAN 2500 OHMS. DURING THE REVISION PROCEDURE, THE RV LEAD WAS NOTED TO BE FRACTURED. THE LEAD WAS SURGICALLY ABANDONED AND A NEW LEAD WAS SUCCESSFULLY IMPLANTED. THE PATIENT IS NOT PACEMAKER DEPENDENT AND EXPERIENCED NO ADVERSE EFFECTS AS A RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FINELINE II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4457 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Hospitalization| L| R | 1297| 4457| 4470 |