FDA Adverse Event Injury Summary report: N

FINELINE II

MDR report key: 1946094 · Received January 4, 2011

Report

Report Number
2124215-2010-20312
Event Type
Injury
Date Received
January 4, 2011
Date of Event
October 14, 2010
Report Date
October 14, 2010
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
PMA / PMN Number
P960004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THIS LEAD WAS SURGICALLY ABANDONED AND A NEW LEAD WAS SUCCESSFULLY IMPLANTED. THERE IS NO FURTHER INFORMATION AVAILABLE AT THIS TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE TO OUR COMPANY, THIS EVENT WOULD BE UPDATED. THE INVESTIGATION IS COMPLETE AT THIS TIME.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED IMPEDANCE MEASUREMENTS GREATER THAN 2500 OHMS. DURING THE REVISION PROCEDURE, THE RV LEAD WAS NOTED TO BE FRACTURED. THE LEAD WAS SURGICALLY ABANDONED AND A NEW LEAD WAS SUCCESSFULLY IMPLANTED. THE PATIENT IS NOT PACEMAKER DEPENDENT AND EXPERIENCED NO ADVERSE EFFECTS AS A RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4457

Patients

Seq Age Sex Outcome Treatment
1 85 YR Hospitalization| L| R 1297| 4457| 4470