FDA Adverse Event Injury Summary report: N

TELIGEN

MDR report key: 1946044 · Received January 4, 2011

Report

Report Number
2124215-2010-20115
Event Type
Injury
Date Received
January 4, 2011
Date of Event
October 14, 2010
Report Date
November 11, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
PMA / PMN Number
P010012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT WILL BE FURTHER UPDATED FOLLOWING COMPLETION OF LAB ANALYSIS. AT THIS TIME, NO ADDITIONAL INFORMATION IS AVAILABLE.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED DUE TO SYSTEM EROSION. NO ADVERSE PATIENT EFFECTS REPORTED AND THE EXPLANTED DEVICE IS EXPECTED TO BE RETURNED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND E102

Patients

Seq Age Sex Outcome Treatment
1 47 YR Required Intervention 0185| E102