FDA Adverse Event Malfunction Summary report: N

ADVANTA

MDR report key: 1945949 · Received December 28, 2010

Report

Report Number
1824206-2010-11935
Event Type
Malfunction
Date Received
December 28, 2010
Date of Event
December 7, 2010
Report Date
December 7, 2010
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ACCOUNT REPLACED THE COMPOSER INTERFACE BOARD TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

ACCOUNT ALLEGED THAT THE COMPOSER INTERFACE BOARD IS CORRODED DUE TO FLUID INGRESS. ACCOUNT STATED THAT THERE WERE NO INJURIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1600

Patients

Seq Age Sex Outcome Treatment
1