FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 1945945
·
Received December 28, 2010
Report
- Report Number
- 1824206-2010-11963
- Event Type
- Malfunction
- Date Received
- December 28, 2010
- Date of Event
- December 10, 2010
- Report Date
- December 10, 2010
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE MANUAL CPR VALVE WAS LEAKING CAUSING THE HEAD CYLINDER TO SLOWLY LOSE PRESSURE. HE REPLACED THE MANUAL CPR VALVE TO REPAIR THE BED.
Description of Event or Problem · 1
INFO REC'D INDICATES THE HEAD OF THE BED WILL NOT STAY UP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |