FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1945945 · Received December 28, 2010

Report

Report Number
1824206-2010-11963
Event Type
Malfunction
Date Received
December 28, 2010
Date of Event
December 10, 2010
Report Date
December 10, 2010
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE MANUAL CPR VALVE WAS LEAKING CAUSING THE HEAD CYLINDER TO SLOWLY LOSE PRESSURE. HE REPLACED THE MANUAL CPR VALVE TO REPAIR THE BED.

Description of Event or Problem · 1

INFO REC'D INDICATES THE HEAD OF THE BED WILL NOT STAY UP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1900

Patients

Seq Age Sex Outcome Treatment
1