SIZE 10 LEG, GLENOID TARGETER
Report
- Report Number
- 1220246-2024-04833
- Event Type
- Malfunction
- Date Received
- June 4, 2024
- Date of Event
- November 17, 2022
- Report Date
- June 4, 2024
- Manufacturer
- ARTHREX, INC.
- Product Code
- KWS
- UDI-DI
- 00888867315372
- PMA / PMN Number
- K222007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- 003
Narratives
THIS 3500A RECORD IS SUBMITTED TO COMPLY WITH AN FDA 483 INSPECTIONAL OBSERVATION ISSUED TO ARTHREX INC ON MAY 5, 2023. ARTHREX HAS REASSESSED THE REPORTABILITY DECISIONS MADE ON HISTORICAL COMPLAINT RECORDS USING REVISED CRITERION. THIS 3500A DOCUMENT IS A RESULT OF THE REASSESSMENT. COMPLAINT IS CONFIRMED. ONE UNPACKAGED AR-5400-10, BATCH NUMBER 052027, WAS RECEIVED FOR INVESTIGATION. FUNCTIONAL TESTING FOUND THAT THE GLENOID TARGETED LEG GETS STUCK WHEN TRYING TO FIT IN THE SLOT. VISUAL EVALUATION FOUND THE DEVICE IS BENT AND HAS MANY GOUGES ALONG THE SHAFT. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO USER ERROR DUE TO USER-APPLIED EXCESSIVE MECHANICAL FORCES. HOWEVER, DUE TO THE DEVICE BEING RETURNED UNPACKAGED, WE ARE UNABLE TO CONFIRM THE REPORTED EVENT.
ON 11/17/2022 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT A AR-5400-10 GLENOID TARGETER THE SIZE 10 LEG JAMMED INTO LETTER D'S SLOT ON THE GLENOID TARGETER SHAFT. OTHER SIZED LEGS WERE ABLE TO SLIDE IN AND OUT FREELY OF LETTER D'S SLOT ON THE GLENOID TARGETER SHAFT. THIS OCCURRED AT THE BEGINNING OF A REVERSE TOTAL SHOULDER ARTHROPLASTY ON (B)(6) 2022 SO THE 11 LEG WAS SUBSTITUTED IN AND THE CASE CARRIED ON SUCCESSFULLY. THERE WAS NO PATIENT EFFECT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 658808 | SIZE 10 LEG, GLENOID TARGETER | PROSTHESIS, SHOULDER, CEMENTED | KWS | ARTHREX, INC. | SIZE 10 LEG, GLENOID TARGETER | 052027 | 00888867315372 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |