FDA Adverse Event Malfunction Summary report: N

R/R ADJ. PIN COLLET 2 - 3.2MM

MDR report key: 1945889 · Received December 28, 2010

Report

Report Number
1811755-2010-02133
Event Type
Malfunction
Date Received
December 28, 2010
Date of Event
December 7, 2010
Report Date
December 7, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
KIJ
PMA / PMN Number
K972367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO ACCEPT THE PIN.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING ROUTINE TESTING AT THE MFR, THE COLLET WOULD NOT ACCEPT THE LARGE PIN. THERE WAS NO PATIENT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 R/R ADJ. PIN COLLET 2 - 3.2MM INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR KIJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK