NONE
Report
- Report Number
- 1219702-2010-00006
- Event Type
- Other
- Date Received
- December 8, 2010
- Product Code
- FPA
- Report Source
- Manufacturer report
Narratives
SET #1 WAS RECEIVED ON (B)(6), 2010. FLUID LEAKAGE WAS CONFIRMED AND THE POINT OF FAILURE WAS VISUALLY IDENTIFIED TO BE ALONG THE JOINT FORMED BETWEEN THE NOZZLE-HEAT EXCHANGER PLENUM. THE PRESENCE OF A SMALL IMPACT DEPRESSION AND INDICATIONS OF COHESIVE FAILURE WITHIN THE JOINT SUPPORTS THE POSSIBILITY OF THE DISPOSABLE SET BEING DAMAGED OUTSIDE OF THE CONFINES OF THE TRAY/POUCH EITHER BY SUSTAINING A MECHANICAL BLOW OR ATTEMPTING INSTALLATION OF THE SET INTO THE PUMP UNIT INCORRECTLY. THERE IS NO EVIDENCE TO SUGGEST THAT THE JOINT FAILURE OCCURRED PRIOR TO OR WHILE IN SERVICE DUE TO ADHESIVE FAILURE AS A RESULT OF EXPOSURE TO MOISTURE AND/OR OTHER ENVIRONMENTAL FACTORS. THE MAGNITUDE OF THE LEAK PATH WAS CONFIRMED TO BE WELL WITHIN OUR DETECTABLE LIMITS AS STABLE TEST PRESSURES COULD NOT BE OBTAINED. THERE IS NO EVIDENCE TO SUGGEST THAT THE NOZZLE-PLENUM JOINT FAILURE THAT RESULTED IN FLUID LEAKAGE OCCURRED AS A RESULT OF WORKMANSHIP, DESIGN, MATERIALS, PRODUCTION PRACTICES, OR NEGLIGENCE BY BELMONT INSTRUMENT. SET #2, RECEIVED ON (B)(4), 2010: FLUID LEAKAGE WAS CONFIRMED AND THE POINT OF FAILURE WAS VISUALLY IDENTIFIED TO BE ALONG THE OUTER CIRCUMFERENTIAL WELD BETWEEN THE HEAT EXCHANGER ASSEMBLY TOP AND BOTTOM COMPONENTS. THE CRACK, ORIGINATING FROM A POINT OF HIGHLY CONCENTRATED MECHANICAL STRESS, ATYPICAL OF A MFG ARTIFACT, WITH ADDITIONAL CRACKS PROPAGATING PERPENDICULAR THROUGH THE WELD AS WELL. THIS HIGHLY CONCENTRATED MECHANICAL STRESS INDICATING THE POSSIBILITY OF THE HEAT EXCHANGER ASSEMBLY BEING DAMAGED OUTSIDE THE CONFINES OF THE TRAY/POUCH EITHER BY SUSTAINING AN IMPACT OR DROP, OR EXPOSURE TO A CONCENTRATED COMPRESSIVE FORCE. THE 3-SPIKE SETS, P/N 903-00006, MFG ARE ROUTINELY TESTED 100% BY AIR PRESSURE DECAY, PRIOR TO POUCHING AND STERILIZATION. THERE IS NO EVIDENCE TO SUGGEST THAT THE JOINT FAILURE OCCURRED PRIOR TO, OR WHILE IN SERVICE, DUE TO ADHESIVE FAILURE, AS A RESULT OF EXPOSURE TO MOISTURE AND/OR OTHER ENVIRONMENTAL FACTORS. THE MAGNITUDE OF THE LEAK PATH WAS CONFIRMED TO BE WELL WITHIN DETECTABLE LIMITS AS STABLE TEST PRESSURES COULD NOT BE OBTAINED. THERE IS NO EVIDENCE TO SUGGEST THAT THE HEAT EXCHANGER ASSEMBLY FLUID LEAK OCCURRED AS A RESULT OF WORKMANSHIP, DESIGN, MATERIALS, PRODUCTION PRACTICES, OR NEGLIGENCE BY BELMONT INSTRUMENT. WE SHIPPED MORE THAN 40,000 SETS EACH YR AND HAVE NOT HAD A SIMILAR REPORT OF THIS TYPE OF INCIDENT FROM ANY OTHER HOSP.
.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NONE | FPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |