FDA Adverse Event
Malfunction
Summary report: N
PIN COLLET
MDR report key: 1945845
·
Received December 28, 2010
Report
- Report Number
- 1811755-2010-02126
- Event Type
- Malfunction
- Date Received
- December 28, 2010
- Date of Event
- December 8, 2010
- Report Date
- December 8, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MFR FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE WIRE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE WIRES WERE STICKING DURING A PROCEDURE. THERE WERE NO ALLEGATIONS OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PIN COLLET | INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR | KIJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |