FDA Adverse Event Injury Summary report: N

INSIGNIA

MDR report key: 1945686 · Received January 4, 2011

Report

Report Number
2124215-2010-21412
Event Type
Injury
Date Received
January 4, 2011
Date of Event
October 12, 2010
Report Date
April 11, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Removal / Correction Number
Z-0187-06 THRU Z-0190-06
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MULTIPLE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION FROM THE FIELD, BUT WERE UNSUCCESSFUL. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THIS DEVICE WAS EXPLANTED TWO YEARS AND FIVE MONTHS LATER FOR NORMAL BATTERY DEPLETION (NBD). THE DEVICE WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENT EXPERIENCED A SYNCOPAL EPISODE OF UNSPECIFIED ETIOLOGY. IT WAS REPORTED THAT THE PACEMAKER BATTERY WAS LOW, HOWEVER NO ADDITIONAL INFORMATION COULD BE OBTAINED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INSIGNIA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND 1290

Patients

Seq Age Sex Outcome Treatment
1 57 YR Life Threatening 4035| 1290| 4086