IROX
Report
- Report Number
- 2124215-2010-20842
- Event Type
- Injury
- Date Received
- January 4, 2011
- Date of Event
- September 16, 2010
- Report Date
- October 12, 2010
- Manufacturer
- GUIDANT ANGLETON/ST. PAUL
- Product Code
- NVN
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE LEAD WAS SURGICALLY ABANDONED AND WILL NOT BE RETURNED FOR TESTING. THEREFORE, BOSTON SCIENTIFIC CRM CANNOT CONFIRM THE REPORTED CLINICAL OBSERVATIONS. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFORMATION IS RECEIVED.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT FOLLOWING AN ELECTIVE PROCEDURE TO REPLACE THIS PATIENT'S PACEMAKER, THIS RIGHT VENTRICULAR LEAD WAS EXHIBITING DECREASED IMPEDANCE MEASUREMENTS AND VARYING R-WAVES MEASUREMENTS BETWEEN THE PACING SYSTEM ANALYZER (PSA) AND THE PACEMAKER. THEREFORE, THE POCKET WAS RE-OPENED TO EVALUATE THE LEAD AND AND INSULATION DAMAGE WAS OBSERVED. THE LEAD WAS REMOVED FROM SERVICE AND SUCCESSFULLY REPLACED. A FEW WEEKS FOLLOWING THIS PROCEDURE, THE PATIENT STATED THAT A PNEUMOTHORAX HAD OCCURRED DURING THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IROX | IMPLANTABLE LEAD | NVN | GUIDANT ANGLETON/ST. PAUL | 431-07 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | 294-03| 4457| 431-07| 1284| S603| 438-05| 5076 |