FDA Adverse Event Injury Summary report: N

UNKNOWN DEPUY FEMORAL COMPONENT

MDR report key: 1945623 · Received December 30, 2010

Report

Report Number
1818910-2010-09947
Event Type
Injury
Date Received
December 30, 2010
Date of Event
December 3, 2010
Report Date
December 3, 2010
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
JWH
PMA / PMN Number
K952830
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE SUBMITTED STEM BOLT CONFIRMED FRACTURE. NO MATERIAL DEFECTS WERE OBSERVED ON THE FRACTURE SURFACE. THE LOCATION AND MANNER OF THE STEM BOLT FRACTURE SUGGESTED IT WAS REPEATEDLY UNDER EXCESSIVE BENDING STRESS FOR A PROLONGED TIME, WHICH COULD INITIATE THE FATIGUE AND EVENTUALLY LEAD TO ITS FRACTURE. NO EVIDENCE WAS FOUND SUGGESTING PRODUCT ERROR WAS A CONTRIBUTING FACTOR AND THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Description of Event or Problem · 1

PT REVISED FOR LOOSE FEMORAL/TIBIA COMPONENTS AT CEMENT/IMPLANT AND CEMENT/BONE MANTLES, OSTEOLYSIS AND POLYETHYLENE WEAR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNKNOWN DEPUY FEMORAL COMPONENT PFCSIG JWH DEPUY ORTHOPAEDICS, INC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 59 YR Required Intervention