VITALITY 2
Report
- Report Number
- 2124215-2010-19932
- Event Type
- Malfunction
- Date Received
- January 4, 2011
- Date of Event
- October 13, 2010
- Report Date
- October 22, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-1047-2007 TO Z-1055-20
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- PHYSICIAN
Narratives
THIS PRODUCT IS CURRENTLY UNDERGOING DETAILED ANALYSIS. THIS EVENT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION AND WAS RETURNED FOR DISPOSAL. ROUTINE INITIAL DEVICE TESTING INDICATED THAT DETAILED ANALYSIS WAS REQUIRED. TO DATE, THERE HAVE BEEN NO REPORTED ADVERSE PATIENT EFFECTS RELATED TO THIS CLINICAL OBSERVATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T165 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | 0157| 4086| E110| T165 |