COMFORTDRIVE 200XDR

MDR report key: 1945487 · Received November 30, 2010

Report

Report Number
3003637274-2010-00048
Event Type
Injury
Date Received
November 30, 2010
Date of Event
October 28, 2010
Report Date
November 3, 2010
Manufacturer
KAVO DENTAL GMBH
Product Code
DZA
PMA / PMN Number
K080677
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A PATIENT RECEIVED A VERY SMALL BURN ON CHEEK WHILE DENTIST WAS PERFORMING A STANDARD TREATMENT ON TOOTH #31. NO OINTMENT OR MEDICATION WAS GIVEN TO PATIENT. DURING PRODUCT EVALUATION WAS FOUND THAT THE HEAD HAS A DENT AT 11. THE DENT KEEPS BACK-CAP DEPRESSED CAUSING DRIVE ASSEMBLY TO OVERHEAT DUE TO EXCESSIVE PRESSURE. THE CAUSE FOR OVERHEATING IS THE DENT ON THE HEAD. (B)(4). KAVO DENTAL (B)(4) (THE MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF KAVO DENTAL (B)(4) (THE IMPORTER).

Description of Event or Problem · 1

A PATIENT RECEIVED A BURN ON CHEEK WHILE DENTIST WAS PERFORMING A STANDARD TREATMENT ON TOOTH# 31. NO OINTMENT OR MEDICATION WAS GIVEN TO PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMFORTDRIVE 200XDR DENTAL HANDPIECE DZA KAVO DENTAL GMBH 200XDR

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention