ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1945248
·
Received December 21, 2010
Report
- Report Number
- 2531779-2010-03099
- Event Type
- Malfunction
- Date Received
- December 21, 2010
- Report Date
- November 27, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THERE WAS NO DAMAGE TO KEYPAD. THE 'OK' BUTTON WAS INTERMITTENTLY RESPONSIVE AND DIFFICULT TO PRESS. THE 'CONTRAST/UP/DOWN' WAS RESPONSIVE AND SPRINGS BACK NORMALLY. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER THE KEY CONTACTS.
Description of Event or Problem · 1
THIS COMPLAINT IS BEING REPORTED DUE TO AN ALLEGED KEYPAD MALFUNCTION. THERE WAS NO REPORT OF PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT . REPLACEMENT PRODUCTS WERE SENT TO THE PT.