FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 1945236 · Received December 10, 2010

Report

Report Number
9615739-2010-00296
Event Type
Malfunction
Date Received
December 10, 2010
Date of Event
November 16, 2010
Report Date
November 16, 2010
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE SIDERAIL CABLES, INTER-CABLES, SIDECOM CABLE AND BOTH CAREGIVER SIDERAIL CONTROLS TO REPAIR THE BED.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THE BED DIDN'T FUNCTION. THE TECHNICIAN FOUND THE SIDERAIL CABLES WERE TORN AND EXPOSING BARE WIRING WHICH SHORTED OUT THE SIDERAIL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAREASSIST BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1