CE INFUSOR LV 1.5, 12 PACK
Report
- Report Number
- 6000001-2011-00004
- Event Type
- Malfunction
- Date Received
- January 3, 2011
- Date of Event
- December 2, 2010
- Report Date
- December 9, 2010
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- PMA / PMN Number
- K041738
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHARMACIST
Narratives
(B)(4). DEVICE EVALUATION: THE DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL EXAMINATION WAS PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A RUPTURED RESERVOIR. THE DEVICE IS A SINGLE USE DEVICE AND WAS DISCARDED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4).ADDITIONAL NARRATIVE: THE DEVICE HAS BEEN RECEIVED BY BAXTER FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION AND/OR SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE.
IT WAS REPORTED TO BAXTER (B)(4) THAT THE RESERVOIR OF ONE (1) CE INFUSOR LV 1.5 "BURST " NEAR THE END OF PATIENT THERAPY. THE DEVICE WAS FILLED WITH 5-FLUOROURACIL (2800MG AND 270ML OF 0.9% SODIUM CHLORIDE AND CONNECTED TO THE PATIENT. THERE IS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CE INFUSOR LV 1.5, 12 PACK | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 10J068 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | NORMAL SALINE| 5-FLUOROURACIL |