FDA Adverse Event Malfunction Summary report: N

4 MOTOR ADVANCE BED

MDR report key: 1945145 · Received December 10, 2010

Report

Report Number
1824206-2010-11507
Event Type
Malfunction
Date Received
December 10, 2010
Date of Event
November 16, 2010
Report Date
November 16, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND LOOSE PINS IN THE COMMUNICATION CABLE AND THE CABLE WAS NOT CONNECTED TO THE COMPOSER INTERFACE BOARD AT THE HEAD OF THE BED. THE TECHNICIAN INSTALLED A CABLE SAVER AND CONNECTED THE CABLE TO THE COMPOSER INTERFACE BOARD TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE ROOM LIGHTS CANNOT BE OPERATED FROM BED. THE PATIENT CANNOT PLACE A NURSE CALL USING THE BED CALL SWITCH. PATIENT CANNOT HEAR HUC RESPOND IN THE EXPECTED LOCATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 4 MOTOR ADVANCE BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1155

Patients

Seq Age Sex Outcome Treatment
1 UNK