FINELINE II
Report
- Report Number
- 2124215-2010-19547
- Event Type
- Injury
- Date Received
- January 3, 2011
- Date of Event
- October 1, 2010
- Report Date
- November 16, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
AT THIS TIME, THE LEAD HAS NOT BEEN RETURNED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE TO OUR COMPANY, THIS EVENT WOULD BE UPDATED.
THIS LEAD HAS BEEN RETURNED AND IS CURRENTLY UNDERGOING ANALYSIS. THE EVENT WILL BE UPDATED ONCE ANALYSIS IS COMPLETE.
UPON RECEIPT AT OUR (B)(4) LABORATORY, IT WAS NOTED THAT THE COMPLETE LEAD WAS RETURNED WITH A SET SCREW MARK ON BOTH OF THE TERMINALS AND TISSUE ENTWINED AT THE BASE OF THE HELIX. FURTHER VISUAL INSPECTION NOTED BLOOD AND BODY FLUID IN THE LEAD LUMEN DUE TO MULTIPLE CUTS IN THE INSULATION. DEFORMED CONDUCTOR COILS WERE ALSO OBSERVED AT 120 MM FROM THE TERMINAL PIN, AND ANALYSIS CONCLUDED IT WAS LIKELY DUE TO THE USE OF A GRABBING TOOL AT EXPLANT. ANALYSIS WAS NOT ABLE TO CONFIRM THE FIELD ALLEGATION OF DISLODGEMENT, HOWEVER THEY WERE ABLE TO CONFIRM THE LEAD BODY DAMAGE AND CONTAMINATION.
--
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THE TWO AND A HALF MONTHS POST IMPLANT THE RIGHT ATRIAL (RA) LEAD DISLODGED. WHEN THE LEAD WAS REMOVED FROM THE PATIENT DURING THE REVISION PROCEDURE, THERE APPEARED TO BE DAMAGE ON THE LEAD BODY NEAR THE SUTURE SLEEVE. FURTHER INSPECTION NOTED THAT THERE WAS ALSO INSULATION DAMAGE AND BLOOD IN THE LEAD LUMEN. A NEW RA LEAD WAS SUCCESSFULLY IMPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FINELINE II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4470 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| L| R | (B)(4)| (B)(4)| (B)(4) |