FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 1944956 · Received January 3, 2011

Report

Report Number
2124215-2010-19585
Event Type
Malfunction
Date Received
January 3, 2011
Date of Event
October 5, 2010
Report Date
November 2, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED UPON COMPLETION OF THE ANALYSIS.

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION RECEIVED FROM THE LOCAL SALES REPRESENTATIVE INDICATED THAT A DEVICE REPLACEMENT PROCEDURE WAS SCHEDULED FOR A DATE IN THE NEAR FUTURE. THE DEVICE PLANNED TO BE RETURNED UPON EXPLANT.

Additional Manufacturer Narrative · 1

ATTEMPTS TO RETRIEVE ADDITIONAL INFORMATION HAVE BEEN MADE. AT THIS TIME THERE IS NO ADDITIONAL INFORMATION AVAILABLE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE THE EVENT WILL BE UPDATED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS AT END OF LIFE (EOL) AFTER APPROXIMATELY 59 MONTHS OF IMPLANT. THE MONITORING VOLTAGE WAS 2.68 VOLTS AND THE CHARGE TIME MEASUREMENT WAS 38.4 SECONDS. A BOSTON SCIENTIFIC CRM TECHNICAL SERVICES CONSULTANT RECOMMENDED DEVICE REPLACEMENT. TO DATE, THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T167

Patients

Seq Age Sex Outcome Treatment
1 58 YR Required Intervention T167| 4469| 0184