FDA Adverse Event Injury Summary report: N

EASYTRAK 2

MDR report key: 1944881 · Received January 3, 2011

Report

Report Number
2124215-2010-19426
Event Type
Injury
Date Received
January 3, 2011
Date of Event
October 1, 2010
Report Date
October 1, 2010
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS EVENT WILL BE UPDATED AS NECESSARY.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH OUT OF RANGE IMPEDANCE MEASUREMENTS OF GREATER THAN 3000 OHMS AS WELL AS LOSS OF CAPTURE AT MAX OUTPUTS. THIS RV LEAD WAS FOUND TO BE FRACTURED. TO DATE, THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4543

Patients

Seq Age Sex Outcome Treatment
1 68 YR Life Threatening 4543