VOYAGER NC CORONARY DILATATION CATHETER
Report
- Report Number
- 2024168-2011-00004
- Event Type
- Malfunction
- Date Received
- January 3, 2011
- Date of Event
- December 10, 2010
- Report Date
- December 13, 2010
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- LOX
- PMA / PMN Number
- P810046
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. DILATATION CATHETER: 4.0 X 12 MM POWERED LACROSS; STENT: 4.0 X 12 MM DRIVER. BALLOON MATERIAL RUPTURES CAN BE AFFECTED BY NUMEROUS FACTORS INCLUDING, BUT IS NOT LIMITED TO, BALLOON DAMAGE DURING MANUFACTURING, MATERIALS, HANDLING, LESION CALCIFICATION AND TORTUOSITY, AN INTERACTION WITH A PREVIOUSLY IMPLANTED STENT, INSUFFICIENT PREPARATION PRIOR TO USE OR FROM INTERACTIONS WITH OTHER DEVICES. SINCE THERE WAS NO REPORT OF ANY LEAKS NOTED DURING PREPARATION FOR USE AND THE CATHETER WAS INITIALLY PRESSURIZED ONCE WITHOUT INCIDENT, THIS SUGGESTS THAT THE BALLOON WAS NOT DAMAGED PRIOR TO THE PROCEDURE. ADDITIONALLY, THE LESION WAS REPORTED AS MODERATELY TORTUOUS, HEAVILY CALCIFIED AND 75% STENOSED, WHICH MAY HAVE CONTRIBUTED TO THE REPORTED DIFFICULTY. IF THE BALLOON EXPERIENCES MECHANICAL DAMAGE AT SOME POINT DURING THE PROCEDURE, THIS CAN CAUSE THE MATERIAL TO WEAKEN AND RUPTURE DURING AN ATTEMPT TO INFLATE THE CATHETER. RETURN OF THE VOYAGER NC MAY HAVE FURTHER AIDED THE INVESTIGATION. A REVIEW OF THE FINISHED PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY NONCONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT AND ALL LOT RELEASE TESTING MET MANUFACTURING CRITERIA. BASED ON THE INFORMATION RECEIVED WITH THIS COMPLAINT, THE REPORTED BALLOON RUPTURE APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE AND NOT A PRODUCT QUALITY DEFICIENCY. ALL DILATATION CATHETERS ARE 100% VISUALLY INSPECTED AND LEAK TESTED DURING THE MANUFACTURING PROCESS. A SAMPLING OF UNITS IS ALSO DESTRUCTIVELY TESTED TO VERIFY RBP AND BALLOON INTEGRITY.
IT WAS REPORTED THAT DURING A PROCEDURE OF A MODERATELY TORTUOUS, HEAVILY CALCIFIED, CONCENTRIC MID RIGHT CORONARY ARTERY LESION, THE VOYAGER NC WAS USED FOR PRE-DILATATION; HOWEVER, DURING THE SECOND INFLATION THE BALLOON RUPTURED AT 18 ATMOSPHERE (ATM). THERE WAS NO REPORTED PATIENT EFFECT. A NON-ABBOTT BALLOON WAS USED TO DEPLOY A NON-ABBOTT STENT TO COMPLETE THE PROCEDURE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VOYAGER NC CORONARY DILATATION CATHETER | CORONARY DILATATION CATHETER | LOX | AV-TEMECULA-CT | 9062961 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |