FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 19448551 · Received June 3, 2024

Report

Report Number
9617229-2024-09347
Event Type
Injury
Date Received
June 3, 2024
Date of Event
May 1, 2024
Report Date
October 8, 2024
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED.

Additional Manufacturer Narrative · 0

THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE BAKER GRADE IV.

Additional Manufacturer Narrative · 0

LABORATORY ANALYSIS SUMMARY: THE DEVICE RELATED TO THE REPORTED EVENT OF CAPSULAR CONTRACTURE WAS RECEIVED ON SEPTEMBER 05, 2024 WITH LOT NUMBER 1193136. BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: CAPSULAR CONTRACTURE: UNABLE TO OBSERVE SINCE IT IS NOT RELATED TO THE DEVICE. AS PER THE INVESTIGATION PROCEDURE DEFORMATION WAS COMPLETED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

CORRECTION TO G.3. AWARE DATE OF SUPPLEMENTAL MEDWATCH #EMDR-(B)(4). AWARE DATE SHOULD HAVE BEEN LISTED AS 8/29/2024.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED ¿URGENT BREAST IMPLANT EXCHANGE FOR EARLY CAPSULAR CONTRACTURE BAKER GRADE IV¿. THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE ¿URGENT BREAST IMPLANT EXCHANGE FOR EARLY CAPSULAR CONTRACTURE BAKER GRADE IV¿. DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE ¿URGENT BREAST IMPLANT EXCHANGE FOR EARLY CAPSULAR CONTRACTURE BAKER GRADE IV¿. DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE ¿URGENT BREAST IMPLANT EXCHANGE FOR EARLY CAPSULAR CONTRACTURE BAKER GRADE IV¿. THE DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE ¿URGENT BREAST IMPLANT EXCHANGE FOR EARLY CAPSULAR CONTRACTURE BAKER GRADE IV¿. DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
629693 INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 1193136

Patients

Seq Age Sex Outcome Treatment
1 43 YR Female Required Intervention