FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1944827 · Received December 10, 2010

Report

Report Number
1824206-2010-11523
Event Type
Malfunction
Date Received
December 10, 2010
Date of Event
November 16, 2010
Report Date
November 16, 2010
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE TREND DOWN VALVE IS DEFECTIVE. HE REPLACED THE TREND VALVE ASSEMBLY TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE HEAD HI/LOW FUNCTION IS DRIFTING DOWN VERY SLOW OVER TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1900

Patients

Seq Age Sex Outcome Treatment
1