FDA Adverse Event Injury Summary report: N

EASYTRAK 2

MDR report key: 1944809 · Received January 3, 2011

Report

Report Number
2124215-2010-21364
Event Type
Injury
Date Received
January 3, 2011
Date of Event
October 8, 2010
Report Date
October 8, 2010
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
PMA / PMN Number
P010012
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSPECTION OF THE LEAD IDENTIFIED DRIED BODILY FLUID IN THE LEAD LUMEN AND ON THE TERMINAL PIN. THE CONDUCTOR COILS WERE DEFORMED IN MANY AREAS OF THE LEAD. THE LEAD WAS PUT THROUGH AND FAILED ELECTRICAL CONTINUITY TESTING. THE POLYURETHANE TUBING WAS PUCKERED APPROXIMATELY 243-253 MM FROM THE TERMINAL PIN. THIS APPEARED TO BE THE SUTURE SLEEVE TIE DOWN SITE. THE INNER INSULATION IS WORN/TORN APPROXIMATELY 267-268MM FROM THE TERMINAL PIN. ADDITIONALLY, THE CONDUCTOR COILS WERE FRACTURED APPROXIMATELY 267-268MM FROM THE TERMINAL PIN AND THE LEAD'S POLYURETHANE APPEARED BENT IN THIS AREA. IT WAS CONCLUDED THAT THE CLINICAL ALLEGATION OF LEAD FRACTURE WAS CONFIRMED BY XRAY AND CONTINUITY TESTING. THIS FRACTURE APPEARS TO BE JUST DISTAL OF THE SUTURE SLEEVE TIE DOWN AREA.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION FROM AN IMPLANT FORM THAT THIS DEVICE WAS TAKEN OUT-OF-SERVICE (OOS). THE PATIENT WAS PRESENTED TO THE ELECTROPHYSIOLOGY (EP) LABORATORY FOR A LEAD REVISION PROCEDURE. THE LEFT VENTRICULAR (LV) LEAD WAS EXTRACTED DUE TO CONDUCTOR FRACTURE. THE PHYSICIAN ELECTED TO REPLACE THE DEVICE BECAUSE THE LV RING SET SCREW WAS NOT OPERATING PROPERLY. SUBSEQUENTLY, BOSTON SCIENTIFIC RECEIVED INFORMATION FROM THE LOCAL REPRESENTATIVE THAT THE LV PORT OF THE REPLACEMENT DEVICE HAD BEEN PLUGGED DUE TO THE UNSUCCESSFUL MULTIPLE ATTEMPTS TO IMPLANT TRANSVENOUS LV LEADS DUE TO PATIENT CONDITION. THE PATIENT WAS THEN PRESENTED TO THE OPERATING ROOM AND AN EPICARDIAL LV LEAD WAS SUCCESSFULLY IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4542

Patients

Seq Age Sex Outcome Treatment
1 55 YR Hospitalization| L| R 4542| H220| N119| T175| 4136| 0185| 4548