FDA Adverse Event Injury Summary report: N

VITALITY 2

MDR report key: 1944795 · Received January 3, 2011

Report

Report Number
2124215-2010-21359
Event Type
Injury
Date Received
January 3, 2011
Date of Event
October 7, 2010
Report Date
October 8, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS DEVICE HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). ERI WAS REACHED DUE TO A LONG CHARGE TIME OF 20 SECONDS. IT WAS ALLEGED THE DEVICE HAD DEPLETED PREMATURELY. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T165

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention