FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 1944786 · Received January 3, 2011

Report

Report Number
2124215-2010-19583
Event Type
Malfunction
Date Received
January 3, 2011
Date of Event
October 6, 2010
Report Date
October 18, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Removal / Correction Number
Z-1192-06 THRU Z-1194-06
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY IN PROGRESS. NO ADDITIONAL INFORMATION IS AVAILABLE. SHOULD ANY ADDITIONAL INFORMATION RELATED TO THIS EVENT BECOME AVAILABLE, THIS EVENT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. ALL LOW VOLTAGE POWER SUPPLY MEASUREMENTS WERE CURRENTLY WITHIN AN APPROPRIATE RANGE; HOWEVER, FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION. THIS ISSUE IS DISCUSSED IN THE (B)(4) 2010 PRODUCT PERFORMANCE REPORT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION. ANOTHER BOSTON SCIENTIFIC CRT-D WAS SUCCESSFULLY IMPLANTED. THIS DEVICE IS PART OF THE SHORTENED REPLACEMENT WINDOW ADVISORY. THIS ADVISORY WAS ORIGINALLY COMMUNICATED APRIL 5, 2007. UPON RECEIPT, INITIAL ANALYSIS DETERMINED THAT THE DEVICE DID NOT MEET EXPECTED LONGEVITY PREDICTIONS AS DESCRIBED IN LABELING. NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H175

Patients

Seq Age Sex Outcome Treatment
1 60 YR 4064| 4518| H175| N118| 0185| 1298| 4035