CONTAK RENEWAL
Report
- Report Number
- 2124215-2010-19583
- Event Type
- Malfunction
- Date Received
- January 3, 2011
- Date of Event
- October 6, 2010
- Report Date
- October 18, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Removal / Correction Number
- Z-1192-06 THRU Z-1194-06
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY IN PROGRESS. NO ADDITIONAL INFORMATION IS AVAILABLE. SHOULD ANY ADDITIONAL INFORMATION RELATED TO THIS EVENT BECOME AVAILABLE, THIS EVENT WILL BE UPDATED.
UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. ALL LOW VOLTAGE POWER SUPPLY MEASUREMENTS WERE CURRENTLY WITHIN AN APPROPRIATE RANGE; HOWEVER, FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION. THIS ISSUE IS DISCUSSED IN THE (B)(4) 2010 PRODUCT PERFORMANCE REPORT.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION. ANOTHER BOSTON SCIENTIFIC CRT-D WAS SUCCESSFULLY IMPLANTED. THIS DEVICE IS PART OF THE SHORTENED REPLACEMENT WINDOW ADVISORY. THIS ADVISORY WAS ORIGINALLY COMMUNICATED APRIL 5, 2007. UPON RECEIPT, INITIAL ANALYSIS DETERMINED THAT THE DEVICE DID NOT MEET EXPECTED LONGEVITY PREDICTIONS AS DESCRIBED IN LABELING. NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H175 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | 4064| 4518| H175| N118| 0185| 1298| 4035 |