DEXTRUS 4135
Report
- Report Number
- 1028232-2010-02783
- Event Type
- Injury
- Date Received
- December 16, 2010
- Date of Event
- October 13, 2010
- Report Date
- December 13, 2010
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVZ
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MFG PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MFG PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MFG.
BOSTON SCIENTIFIC CRM RECEIVED INFO THAT THIS DEXTRUS ATRIAL LEAD HAD DISLODGED. THE PT HAD BEEN IN AND OUT OF THE HOSPITAL INTERMITTENTLY OVER THE PAST FEW MONTHS DUE TO CHEST PAIN. THE ASSOCIATED PACEMAKER WAS REPROGRAMMED TO VVI CONFIGURATION. A REVISION PROCEDURE WAS SUBSEQUENTLY PERFORMED AND THE LEAD WAS SUCCESSFULLY REPOSITIONED. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFO IS RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DEXTRUS 4135 | PACER LEAD | NVZ | BIOTRONIK SE & CO. KG | 358744 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |