FDA Adverse Event Injury Summary report: N

DEXTRUS 4137

MDR report key: 1944384 · Received December 16, 2010

Report

Report Number
1028232-2010-02792
Event Type
Injury
Date Received
December 16, 2010
Date of Event
October 29, 2010
Report Date
December 13, 2010
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVZ
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MFG PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MFG PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MFG.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFO THAT THIS PT HAD EXPERIENCED A SYNCOPAL EPISODE AND WAS ADMITTED TO THE HOSPITAL. DEVICE EVAL NOTED THREE SECOND PACING PAUSES AND THIS RIGHT VENTRICULAR DEXTRUS LEAD WAS FOUND TO BE MICRODISLODGED. THEREFORE, A REVISION PROCEDURE WAS PERFORMED, AND THE LEAD WAS SUCCESSFULLY REPOSITIONED. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFO IS RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEXTRUS 4137 PACER LEAD NVZ BIOTRONIK SE & CO. KG 358764

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization