FDA Adverse Event Injury Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1944372 · Received December 21, 2010

Report

Report Number
2531779-2010-03121
Event Type
Injury
Date Received
December 21, 2010
Date of Event
November 27, 2010
Report Date
November 27, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED TO ANIMAS FOR EVAL. IF THE DEVICE IS RETURNED AN EVAL WILL BE CONDUCTED AND THE RESULTS WILL BE PROVIDED IN A SUPPLEMENTAL REPORT. THERE IS NO EVIDENCE OF A PUMP MALFUNCTION AT THIS TIME AND TROUBLESHOOTING DEMONSTRATED THAT THE PUMP WAS DELIVERING INSULIN AS PROGRAMMED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE PT'S MOTHER REPORTED THAT THE PT'S DOCTOR WANTED THE PUMP REPLACED BECAUSE HIS BLOOD GLUCOSE LEVELS WERE RUNNING ABOVE 600 MG/DL. THE MOTHER CONFIRMED THE PRESENCE OF LARGE KETONES. THE PT'S SITE WAS CHANGED BEFORE THE ELEVATED BLOOD GLUCOSE LEVELS OCCURRED AND THE MOTHER NOTED NO ISSUES WITH THE INFUSION SITE. THE PT'S MOTHER CONFIRMED THAT ALL SETTINGS WERE CORRECT IN THE PUMP AND THE DELIVERY HISTORY MATCHED THE PUMP SETTINGS. THE PT'S DOCTOR PRESCRIBED INSULIN INJECTIONS AND THE PT'S MOTHER CONFIRMED THAT THE PT'S BLOOD GLUCOSE NORMALIZED AFTER USING INJECTIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 5 YR Life Threatening| R