FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1944286 · Received December 6, 2010

Report

Report Number
1720753-2010-04452
Event Type
Malfunction
Date Received
December 6, 2010
Date of Event
September 20, 2010
Report Date
December 6, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CANCELED THE SERVICE CALL.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9900 SYSTEM WOULD NOT POWER ON. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1