FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 1944269 · Received December 6, 2010

Report

Report Number
1720753-2010-04489
Event Type
Malfunction
Date Received
December 6, 2010
Date of Event
November 17, 2010
Report Date
December 6, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND CALIBRATED THE COLLIMATOR STOPS. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM DISPLAYED INTERMITTENT COLLIMATOR IRIS ERRORS. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1